Process Equipment Maintenance Series-Single-Use pH Probe Calibration (Part II)

 

describes the use of the probe's own information for slope/intercept value calibration,In order to improve the detection accuracy of the disposable pH probe, after the above calibration is completed, it can be calibrated again as follows, this calibration is an option for Hamilton and MTTLER TOLEDO is a must.

 

 

Process Calibration

 

 

is different from the repeated use of pH probe, because the disposable pH probe has been installed in the disposable bag or pipeline after irradiation sterilization, in order to ensure its sterility, the pH probe cannot be taken out and calibrated with a variety of pH standard solutions,need to use the sample solution for process calibration.

 

disposable pH probe in the state of dry preservation, the life of the probe after the factory 18-36 months.

 

especially important to note that it requires a wetting activation before calibration. Different manufacturers and models require different wetting time. METTLER TOLEDO InSUS series time is 20-120min,Hamilton Onepferm time is 30min, and Finesse Trufluor time is 240-360min.

 

pH probe is complete, perform the process calibration as follows:

 

 

a. Press the calibration button on the transmitter interface, select "Calibrate Sensor", and select the corresponding channel "CHAN_1" or "CHAN_2";

 

 

B. Click "Method" to select "Process";

 

 

 

c. Press the "Cal" button to capture the pH of the sample and press Enter. (At this time, sample solution should be sampled and offline pH detection should be carried out);

 

 

 

d. Enter the pH value of the sample solution in the following interface;

 

 

 

 

e. and press "OK"-"Next"-"SaveCal", the transmitter is calibrated and the calibration data is saved.

 

 

 

 

Single point calibration/Two point calibration

 

 

This method is often used for re-use pH probes, and the operation is similar to the calibration process described above,select "1-Point" or "2-Point" at "Method", the sample solution is replaced with the pH standard solution, and the subsequent operation is the same as the above process calibration, which will not be repeated in this article.

 

teachers, if there are any details in this article, please contact RealTop Technical Service Hotline 0513-82104020 or send an email to service@realtopbio.com for our technical support.

 

 

cooperation

 

support

 

 

welcome teachers in need to inquire!

 

RealTop Biology can provide a disposable mixing system with a volume of 10L-3000L, and can be equipped with online detection functions such as weighing, jacket, temperature, pH, conductivity, pressure, etc,welcome teachers in need to inquire!

 

 

 

 

 

 

 

Created on:2023-02-08 17:00
Home Page    Product Information    Process Equipment Maintenance Series-Single-Use pH Probe Calibration (Part II)
PV:0
  • 回到顶部
  • 0513-82104020
  • QQ客服
  • Wechat QR code